Page 12 - qp_17
P. 12
SENSITIVITY AND SPECIFICITY STUDY
PROCEDURE
ANALYTE / TEST : PanBio Dengue IgM Capture ELISA
METHOD : Enzyme-linked Imunnoassay
ANALYZER : Microplate Reader (Biorad)
Known Condition
Positive Negative
(No. of samples = 20) (No. of samples = 20)
Positive TP =18 FP =1 PPV = 95%
Test
outcome
Negative FN =2 TN =19 NPV = 95%
Sensitivity Specificity
= 95 % = 95%
False positive rate (%) = 1 – specificity (95)
= 5%
False negative rate (%) =1 - sensitivity
= 5%